With the increasing complexity of ICs, test for electromagnetic
radiation and susceptibility becomes an urgent requirement to meet new
demands in design. Requirements for low EMI
and strong anti-EMI
capacity demand further improvements on electronic devices and systems
to meet EMC standards.
Moreover, product engineers must address EMC issues, particularly
meeting threshold requirements and reduced EMI between the systems and
other electronic devices. IC integration density doubles almost every
year, making EMC a key priority in design.
To address the issue, the International Electrotechnical Commission
(IEC) released standards IEC 61967, for EME test of ICs with
frequency
from 150kHz to 1GHz, and IEC 62132 for EMS test of ICs with frequency
from 150kHz to1GHz. Scheduled to be released soon is IEC 62215, a
complementary standard to IEC 62132 that addresses EM disturbance on
ICs.
EME testing
IEC 61967 requires the use of a transverse
electromagnetic cell (TEM
cell) and field probe to measure electromagnetic
emission (EME). It
uses spectrometers or receivers with frequency range from 150kHz to
1GHz. Normally, the ICs to be tested are placed on one circuit board.
The specification of the circuit board is shown in Figure 1.
 |
| Figure
1. ICs to be tested using TEM cell method are placed on one circuit
board. |
A power supply with low RF noise is needed for testing while the
temperature should be at 23°C (±5°C). There can be cases
of environmental noise as peripheral circuits are measured. If the
background noise is lower than the threshold value by more than 6dB, a
preamplifier is applied, if needed.
This method measures radiation using a TEM cell, which is an
irregular coaxial line. In the middle of the line, a flat chip such as
an inner conductor with a quadrate outer conductor connects to the
receiver at one end and the matched load at the other end (Figure 2).
 |
| Figure
2. A flat chip as an inner conductor with a quadrate outer conductor
connects to the receiver at one end and the matched load at the other
end. |
A quadrate vent at the top of the outer conductor is placed to fix
the circuit board to be tested. Specifically, the IC will be installed
inside the cell, with the wiring side and the peripheral circuits
facing outside. This way, the radiation emission will mainly come from
the IC being tested.
High-frequency current produced by the tested chip will flow along
the wire, while wielding pins and package connections serve as the
antenna of radiation emission. If the testing frequency is lower than
the first-order high-mode frequency, only TEM as the major mode is
transmitted. Note that the power of the TEM cell's port has a
quantitative relation with the emission power of the disturbance
source.
Hence, the voltage value collected at a curtain area of the TEM cell
can help evaluate the IC's radiation emission.
 |
| Figure
3. The surface scan method tests the spatial distribution of the IC's
surface electric and magnetic fields. |
The TEM cell method implements surface scan by testing the spatial
distribution of the IC's surface electric field and magnetic field. In Figure 3, the electronic or the
magnetic field probe scans over the IC surface.
The frequency, emission value and spatial position are collected
every time and processed by the software. Spatial frequency
distribution of the field intensity is displayed in color figures. The
result's quality relates closely to the precision of the mechanical
locating system and the size of the probes.
Hence, this is applicable to general PCBs and may not necessarily
use the standard testing circuit boards recommended.
By scanning the electronic magnetic fields over the IC surface, we
can get information about the relative intensity of the EM radiation
source. This method can accurately identify the area on the IC that has
too much radiation.
Partially-shielded electronic field probes and single-loop
minielectronic field probes are recommended in this standard. Both
probes can be made by 0.5mm semi-rigid coaxial cables.
Radiation measurement
The standard has three methods for radiation measurement: direct
coupling, workbench faraday cage (WFC) and magnetic field probe.
There are two measurement types for the direct coupling method: 1
ohm and 15 ohm. The 1 ohm method tests the total disturbance current of
grounded pins, while the 15 ohm tests the disturbance voltage of output
ports.
After going through the chip, RF current will flow to the IC's
grounded pins, so the measurement against the ground RF current can
reflect more precisely the IC electromagnetic disturbance. If you
connect a 1 ohm resistance in serial in the ground loop, you can obtain
the RF current from the loop. Meanwhile, measure the impedance matching
of the testing instruments and the grounding pin.
 |
| Figure
4. The 150ohm direct coupling method tests the disturbance voltage of
single- or multi-output signal lines for impedance matching. |
The 150 ohm method can be adopted to test the disturbance voltage of
single- or multi-out-put signal lines (Figure
4). Note that 150 ohm is the statistic average value of
wire-harness common impedance. To match it with a 50 ohm impedance
testing system, apply a matching network.
Meanwhile, the WFC method of the standard measures the conduction
disturbance voltage of power lines and I/O signal lines. Place a
standard or an application circuit board with the IC to be tested into
the WFC.
 |
| Table
1: EMC testing also requires selection of resolution bandwidth for
spectrum analyzers and receivers. |
All the power and signal lines in and out of the WFC will be
filtered. Connect the WFC testing ground port to testing instruments
and the ground port to be tested to a 50 ohm matching load. A
well-shielded environment reduces the background noise. You can connect
a 100 ohm resistance in serial to match the impedance (Figure 5).
 |
| Figure
5. The IEC 61967 WFC method measures the conduction disturbance voltage
of power lines and I/O signal lines. |
Magnetic probe
The magnetic probe method of the standard evaluates the radiation
emission of ICs by testing the current through PCB wires (Figure 6).
The pins of the chip are connected to the power source or periphery
circuits via wires on the PCB; thus, the magnetic probe can obtain the
RF current.
 |
| Figure
6. The magnetic probe method evaluates the radiation emission of ICs by
testing the current through PCB wires. |
Based on the law of electromagnetic
induction, the voltage at the
probe's output end is proportional to the RF current of the wires. An
easy and repeatable method is needed to evaluate the interference
immunity of a chip, which is divided into two fields - radiation
immunity and conduction immunity.
This method collects the RF power if ICs fail to work. IEC 62132
divides the operation status of an IC into five grades; it will then
detect the continuous and AM waves separately. Similar to ISO11452,
peak-level constant AM with depth of 80 percent 1kHz is applied.
 |
| Figure
7. The TEM cell method can also be used for susceptibility test. |
This is different from IEC 61000-4-3. The IEC standard demands that
the peak power of the modulation signal be 1.8x the carrier wave
signal. However, when the peak level constant is under testing, the
modulated signal power, with a modulation depth of 80 percent, is only
0.407x the carrier wave signal power.
The TEM cell discussed in IEC 61967 is also suitable for radiated
susceptibility measurement. One end of the cell is connected to a
signal generator and PA, while the other end is connected to a proper
matching load. The TEM waves - produced in the cell and similar to
remote TEM waves - can be used to measure EMS. Other status monitoring
instruments may be required to monitor the IC status in real-time.
 |
| Figure
8. In the BCI method, interference power is injected into the single
wire or wire harness of IC pins. |
Set of testing
The standard uses three methods for the susceptibility test: bulk
current injection (BCI), direct power injection (DPI) and WFC. In the
BCI method, interference power is injected into the single wire or wire
harness of IC pins. Because of inductive coupling, the cable tested
produces the disturbance current that can be detected by another
current probe. The method derives from the automobile electronic
susceptibility test.
 |
| Figure
9. DPI test injects capacitive coupling to measure susceptibility. |
Compared to BCI that injects inductive coupling, DPI injects
capacitive coupling. RF signals are injected directly to a single pin
or a group of pins of the chip. The coupling capacitance prevents DC
from being directly placed on the output end of AP (Figure 9). The WFC
method in IEC 61967 can also be applied for susceptibility test, but
the receiver is replaced with signal resource and AP (Figure 10).
 |
| Figure
10. The WFC method in IEC 61967 can also test susceptibility, with the
receiver replaced with signal resource and AP |
The shielded structure and good filtering hold RF disturbance
signals inside the cage, thus protecting operation personnel.
Ding Ding is a Product Engineer
at Rohde-Schwarz
China Ltd
|